MRC:
A recording of the physical,functional,and performance characteristics for an item of supply.TEXT
General Description:
Can test the oximeter w/o the probe in-line; probe analysis includes: led test - voltage drop across both red and infrared diodes, photodiode text detector response to a 1ma illuminating current and resistance between all wires; can simulate arterial oxygen saturation from 35-100 pct oxygenation; simuulates patient conditions such as normal, weak pulse, geriatric, obese, bradycardia and tachycardia; portable; 9 preload manufacturers r-curves; 5 downloadale r-curves; physiological finger for sp02 test; rs232 port; physiological motion artifact; heartrate range: 30-250 bpm: resolution: 1 bpm; pulse amplitude range: 0-20 pct perfusion, resolution: 1 pct; 2-line x 24-character super twist lcd display