MRC:
The name and number of parts supplied with the item which may be required for application.AKYD
Accessory Components And Quantity:
Reusable mask: 1 adult, 1 pediatric, and 1 infant. Reusable breathing circuit: 1 adult and 1 pediatric. User manuals, 2 ea.; maintenance manual, 2 ea. Reuable container: 1 ea.
MRC:
Those unusual or unique characteristics or qualities of an item not covered in the other requirements and which are determined to be essential for identification.FEAT
Special Features:
A portable field anesthesia machine (fam), permitting use of closed (low flow), semi-closed and opened anesthesia techniques. 110/220 vac, 50/60 hz. With battery backup to fully power all operations for a of one hour. The unit c/o: a carbon dioxide (co2) absorber, oxygen (o2) absorber, ventilator, and vaporizer. Has nitrous oxide capability and has a operational range of 40-61 psi. Accepts an oxygen concentrator (when available), air compressor, oxygen cylinders, pipeline oxygen, nitrous oxide, and other air sources. Unit is adaptable to use both active and passive gas scavenging systems and is supplied with gas pressures guages, break-resistant flowmeters, and an oxygen flush valve. Monitoring capabilities: fraction of inspired oxygen (fio2), peak airway pressure (paw), and exhaled volume. The vaporizer has a min. Cap. Of 100 ml and the ventilator has a flow rate of 0-50 lpm. The unit is sterilizable iaw with depmeds procedures.