부품 정보
Tube unit medical xray apparatus focal spot size 0.3; 1.0mm sq
TXT
묘사:
Medical X-ray Apparatu Tube Unit
MRC:
Indicates the type of container provided.ALQG
Container Type:
Any acceptable
MRC:
Indicates the type of tube insert anode provided with the item.CLPJ
Tube Insert Anode Type:
High speed rotating
MRC:
The maximum instantaneous voltage for which the radiographic tube insert is rated.CLPL
Tube Insert Radiographic Peak Voltage Rating:
150.0 kilovolts
MRC:
The number of focal spots provided.CLPP
Focal Spot Quantity:
2
MRC:
Designates the size of the focal spot.CLPQ
Focal Spot Size:
0.3 mm sq; @1.0 mm sq
MRC:
An indication of the x-ray protective value of the housing,excluding port.CMGS
Housing X-ray Protective Value:
X-ray protective value of housing shall be in accordance with dhew publication no. Fda 75-8003 jul 1974
MRC:
The element,compound,or mixture of which the target base is fabricated.CQTT
Target Base Material:
Molybdenum
MRC:
The angle of the target in relation to the path of electron flow,expressed in degrees.CQWH
Target Angle In Deg:
10.0
MRC:
The element,compound,or mixture of which the target face is fabricated.CSJQ
Target Face Material:
Tungsten-rhenium alloy
MRC:
The angle of the cable receptacle in relation to the port,expressed in degrees.CTBN
Port To Cable Receptacle Angle In Deg:
270.0
MRC:
The heat capacity,in heat units,for which the anode is rated.CTGC
Anode Heat Capacity Rating In Heat Units:
300000