부품 정보
Standard platinum resistance thermometer; meyers thermometer
TXT
묘사:
Electric Temperature Transmitter
MRC:
The dimension measured along the longitudinal axis with terminated points at the extreme ends of the item.ABHP
Overall Length:
470.0 millimeters
MRC:
The length of a straight line which passes through the center of a circular figure or body,and terminates at the outside circumference.ABKV
Outside Diameter:
7.5 millimeters
MRC:
The minimum and maximum limits of temperature at which the item is rated for operation.AFGA
Operating Tempurature Range:
-200.0/+500.0 degrees celsius
MRC:
Indicates the type of unit.APQB
Unit Type:
Protective tube
MRC:
Indicates the type of case provided.BHMC
Case Type:
Hinged lid carrying case w/handle
MRC:
Indicates the type and number of contacts accommodated by the item.CTXB
Accommodated Contact Type And Quantity:
4 lug
MRC:
Those unusual or unique characteristics or qualities of an item not covered in the other requirements and which are determined to be essential for identification.FEAT
Special Features:
Length 470mm excluding handle; silver plated copper leads 2.4m long; immersion depth 7 to 14 inches; sesnsing length 33mm; time constant 6 seconds; resistance at 0 degrees c is 25.5 ohms' sheath material is quartz, filled with dry air; accuarcy is within 1 micro kelvin
MRC:
Describes the capabilities,intended use,and/or purpose for which the item is provided.FTLD
Functional Description:
Used to measure the temperature of triple point and freezing point cells used in the calibration of thermocouples and other types of thermometers
MRC:
An indication of the precious material and its location in the item.PMLC
Precious Material And Location:
Terminals gold and resistor and internal leads to handle platinum and silver plated copper lead wires silver
MRC:
Identification of the precious material contained in the item.PRMT
Precious Material:
Gold and platinum and silver
MRC:
Literature and references available for information pertaining to the item.RDAL
Reference Data And Literature:
Mfg commercial catalog
MRC:
Indicates the type and number of terminals for providing electrical connection.TTQY
Terminal Type And Quantity:
4 lug