TXT
묘사:
Environmental Testing Chamber
MRC:
Indicates the type of current whether alternating,direct,or both.ACDC
Current Type:
Ac
MRC:
The number of complete cyclic changes,per unit of time,for which an item is rated.ACZB
Frequency Rating:
60.0 hertz
MRC:
An outside measurement taken at right angles to the length of an item,in distinction from thickness.ADJV
Outside Width:
60.000 inches
MRC:
An outside measurement from the bottom to the top of an item,in distinction from depth.AFMW
Outside Height:
72.000 inches
MRC:
The value(s) of potential for which the item is rated.AMSE
Voltage Rating:
220.0 volts and 440.0 volts
MRC:
An outside measurement between specified points on an item,in distinction from height.BCXB
Outside Depth:
96.000 inches
MRC:
The means by which the item is heated.BDWT
Heating Method:
Electrical
MRC:
Indicates the type of environmental chamber.CWZY
Environmental Chamber Type:
Humidity and altitude and temperature
MRC:
The number of alternating current phases.FAAZ
Phase:
Three
MRC:
The chemical compound or mechanical mixture properties of which the item is fabricated.MATT
Material:
Steel
MRC:
Characteristics or qualities of an item,not covered in any other requirement,which are considered essential information for one or more functions excluding nsn assignment.SUPP
Supplementary Features:
Refrig capable lowering temp p75 to m100 degrees f in 160 min; heating sys capable of heating from p75 to p350 degrees f in 140 min; vacuum sys capable of pumping chamber from ambient to 100000ft altitude in 35 min; humidity sys capable of creating 20 to 95 pct rh in range 35 to 185 degrees f; remote control console w/bristol series 500 2 case program/recorders