부품 정보
Mechanical convection oven
MRC:
The distance measured in a straight line from the bottom to the top of an item.ABKW
Overall Height:
32.000 inches
MRC:
An overall measurement taken at right angles to the length of an item,in distinction from thickness.ABMK
Overall Width:
24.000 inches
MRC:
Indicates the type of current whether alternating,direct,or both.ACDC
Current Type:
Ac
MRC:
The number of complete cyclic changes,per unit of time,for which an item is rated.ACZB
Frequency Rating:
Between 50.0 hertz and 60.0 hertz
MRC:
An inside measurement taken at right angles to the length of an item,in distinction from thickness.ADJT
Inside Width:
23.000 inches
MRC:
An inside measurement between specified points of an item,in distinction from height.AFEF
Inside Depth:
19.000 inches
MRC:
An inside measurement from the bottom to the top of an item,in distinction from depth.AFMQ
Inside Height:
19.000 inches
MRC:
The value(s) of potential for which the item is rated.AMSE
Voltage Rating:
120.0 volts
MRC:
The amount of electrical energy that can be dissipated.ANPT
Power Rating:
1735.0 watts
MRC:
Indicates the type of control.AQZF
Control Type:
Temperature
MRC:
The minimum and maximum operating temperatures at which the item is rated,expressed in degrees celsius.ASQJ
Operating Tempurature Range In Deg Celsius:
+35.0/+225.0
MRC:
The heat source for which the item is designed.AZZH
Heat Source For Which Designed:
Integral
MRC:
The means by which the item is heated.BDWT
Heating Method:
Electrical
MRC:
An indication of whether or not the item is provided with a water jacket.BGXH
Water Jacket:
Not provided
MRC:
The means used to provide circulation.BHGT
Circulation Method:
Mechanical
MRC:
An indication of whether or not an insulated cabinet is provided.BLST
Insulated Cabinet:
Provided
MRC:
The element,compound,or mixture of which the cabinet outer wall is fabricated,excluding any surface treatment.BQMH
Cabinet Outer Wall Material:
Steel
MRC:
The element,compound,or mixture of which the cabinet inner wall is fabricated,excluding any surface treatment.BQMJ
Cabinet Inner Wall Material:
Steel, corrosion resisting
MRC:
The number of outer doors included.BQMN
Outer Door Quantity:
1
MRC:
The element,compound,or mixture of which the outer door is fabricated,excluding any surface treatment.BQMP
Outer Door Material:
Steel
MRC:
An indication of whether or not an outer door insulation feature is included.BQMR
Outer Door Insulation Feature:
Included
MRC:
The number of alternating current phases.FAAZ
Phase:
Single
MRC:
Those unusual or unique characteristics or qualities of an item not covered in the other requirements and which are determined to be essential for identification.FEAT
Special Features:
Sensitivity +/- 0.25 degrees c; includes adjustable nickel plated steel wire shelves; magnetic door closure; adjustable air exhaust; exterior finished in baked silicone enamel; insulated on sides with 2.500 inches of fiberglass; single knob dual hydraulic thermostat-one for control, one for automatic safety to prevent temperature overshooting; recovery time 18 minutes